This guide details the process for filtration using filter agents using vacuum filtration with a sinter funnel. The pictures below show both Celite and silica gel being employed as filter agents.
The filter agent can be prepared a little in advance, but care needs to be taken to ensure that the filter agent pad is not allowed to dry out before use.
Set up a Buchner flask, Buchner ring and sinter funnel following the guidance for Sinter Filtration. The pictures in this guide assume a Buchner flask is being employed, but the steps are the same for filter adapters with a round bottomed flask.
A sinter funnel setup in a fume hood with the Buchner flask attached to the vacuum line.
The relevant filter agent should be added into the sinter funnel. The amount of filter agent may be specified in the procedure, this might be:
As a mass - in which case it can be pre-weighed in a beaker (usually within a fume hood)
As a depth once packed - as a rule of thumb most filter agents pack down to about half the depth once solvated
There are a number of ways to add the filter agent. Some common routes include:
Using the sinter funnel to 'scoop up' the material
Using a beaker to transfer material into the funnel
Via spatula or spoon (note this is likely to be very slow)
Many filter agents present respiritory hazards due to the fine nature of the particles. Care should be taken to ensure that the material is contained within fume hoods during use, and any spills are cleaned up promptly. In most cases damp paper towel should be used to collect any spilled material, taking care not to aerosolise any material. Seek advice from demonstrators as required.
The choice of solvent is usually determined by the mixture being filtered. In most cases the filter agent will be prepared using the same solvent as the material to be filtered is in.
It is also possible to use solvent gradients during the filtration process, but changing the solvent composition during the filtration stages. Where this is desired, the filter agent should be prepared with the first solvent for the gradient process.
Some commonly employed solvents for active filtration include:
Ethyl acetate
Cyclohexane
Ethanol
Ethyl acetate
Acetone
Water
Or mixtures, particularly ethyl acetate/cyclohexane (or petroleum ether) mixes.
Using a spatula, the filter agent is thoroughly stirred into the solvent in order to displace any air bubbles. Care should be taken to ensure that all the solid material is fully suspended in the solvent.
One a homegenus suspension has been obtained, the vacuum can be turned on to cause the filter agent to compact into a 'pad'. Ideally the vacuum should be released as soon as the solvent level reaches the top of the pad, so that the pad remains solvated.
If the pad is not of sufficient depth when formed, then additional filter agent can be added to the sinter, followed by solvent. The entire mixture should be resuspended in the solvent and the pad reformed. This process can be repeated until the pad meets requirements.
Before filtering any reaction mixtures, the Buchner flask should be emptied of any solvent used to prepare the filter agent.
The mixture to be filtered should be carefully poured onto the surface of the pad. Care should be taken to ideally:
Puring gently to avoid disturbing the surface of the pad
Pouring onto the middle of the pad, avoiding the edges, so that materal doesn't travel down the glass sides of the sinter rather than through the material.
The filter agent pad should be washed through with sufficient portions of additional solvent to ensure that any desired material has been eluted from the pad into the Buchner flask as filtrate. Washings may wish to include the vessel from which the material to be filtered was contained in.
If only a single solvent system is being used, then in most cases the solvent pad may be allowed to suck dry with all the solvent passing into the Buchner flask as filtrate.
Ensure that the level of solvent in the Buchner flask is not allowed to exceed the working capacity of the flask.
Once the filtration is complete, the vacuum can be disconnected and the sinter funnel removed. The filtrate from the Buchner flask can then be collected for any further steps.
In order to dispose of the filter agent and any collected residues or absorbed compound, the pad will need to be broken up using a spatula. The material should be carefully broken up so that it can be transferred to an appropriate waste container for this material. Care should be taken whilst breaking up the material to avoid damaging the sintered frit, which can be easily damaged if scraped with spatulas.
Considerations should be given to the hazards of the materials involved. There may be specific waste disposal requirements for one or more components.
Once the solid material has been removed, usually any remaining traces can be removed from the sinter using water at the sinks. The solid material will not pass through the sinter, so needs to be removed carefully from the glass and surface of the sinter. Once any traces of filter agent have been removed, the sinter can be washed as normal.
Usually the final rinse through should be using acetone on a Buchner flask setup to clean the sintered frit. For more information see the cleaning glassware pages.