Stuart SMP30

Instrument

Photo of a stuart SMP30 melting point instrument

Specification

Melting point range

Ambient to 400 °C

Ramp rate 

0.55 to 10.0 °C min-1 

Accuracy

50 °C: ± 0.5 °C

100 °C: ± 1.0 °C

360 °C: ± 2.5 °C

Resolution

0.1 °C

Heating time

350 °C in 6 min

Cooling time

350 °C to 50 °C in 12 min

Detector

PT1000 Platinum resistance temperature detector (Pt-RTD)

Instructions  

Students

Stuart SMP30 Instructions (PDF) ↓

Videos

Stuart SMP30 Melting Point Apparatus