Jasco
Instrument
Teaching Labs had four Jasco HPLC systems consisting of PU-1585 pumps with UV-975 detectors these have now been replaced by newer Agilent models.
Pump Specification
Model
Jasco PU-1585
Pump type
Isocratic dual piston
Injection loop volume
20 μL
Flow range
Detector Specification
Model
Jasco UV-975
Wavelength range
190-600 nm
Wavelength accuracy
±2 nm
Light source
Deuterium lamp
Flow cell volume
8 μL
Software
Clarity
Accessing data
Data from the HPLC instruments is automatically made available at ftp://chemtl.york.ac.uk/HPLC-Jasco/.
Data is autosaved with a filename based on the instrument name, date and time of injection (InstrumentName - dd_mm_yyyy hh_mm_ss). Data is available in the following formats:
PDF
Portable document format. An auto processed document containing the chromatogram and automatically generated peak table with integrations.
CDF
Chromatography Data Interchange Format. This format can be imported directly into ACD/Spectrus for reprocessing.
TXT
Tab-delimited text format with time in minutes against voltage in millivolts. This can be read into any application capable of processing text data (ex Microsoft Excel).
PRM
Native chromatography file for the Clarity which controls the HPLCs.