Modified Atmosphere Gas Flows

Setup

This section will explain how to set up a reaction where the solvent is degassed using a Pasteur pipette bubbler with a nitrogen flow, before removing the solvent degassing setup and switching to maintaining the reaction under a nitrogen atmosphere with a bubbler. In this setup, a 100 mL two-neck round bottomed flask (with stirrer bar) and condenser will be assembled, the solvent degassed, and the reaction setup held under nitrogen using the nitrogen flow attached to the bubbler on the top of the condenser.


This is a fairly common approach to using this setup, although other methods might be encountered. It is worth noting that it is uncommon to maintain a Pasteur pipette bubber during a reflux, as the flow of nitrogen is likely to result in the loss of solvent from the reaction. Hence, it is usual to degas the solvent first, then maintain the reaction under a flow of nitrogen from the top of the condenser, which minimises solvent loss from the setup.

Pasteur pipette fitting

A thermometer adapter is used to hold a Pasteur pipette in place, which needs to be adjusted to be a suitable length to reach below the surface of the solvent in the flask. Care needs to be taken to ensure the pipette doesn't get damaged by the stirrer bar in the flask. It is likely some trial and error adjustment will need to be made, and it is sensible to do this prior to adding any solvent or reagents into the reaction setup.


To fit a Pasteur pipette into a thermometer adapter you will need to carry out the following steps. Care should be taken whilst fitting the pipette, as these are particularly fragile, and may break which could result in injury. Care should also be taken when fitting or removing tubing to the pipette. Note: only low pressure gas sources should be used, where the pressure control is adequate for the reaction setup. The nitrogen supply is our teaching labs is suitable, the compressed air supply must not be used as this is at too high a pressure.



At this stage, solvent could be introduced and a gas bubbled through the solvent without a bubbler being fitted. This is sometimes used for reagent gases (eg carbon dioxide, hydrogen or oxygen), but it is more common to fit the bubber first.

Dismantled thermometer adapter

The plastic screw cap, glass cone and screw thread, and the rubber seal.

Assembed Pasteur pipette in a thermometer adapter

Note the addition of Parafilm® to enable a secure fit onto tubing.

Bubbler fitting

A glass bubbler is used to sit on top of the reaction reflux condenser. The bulb of the bubbler needs to be half filled to form a seal, usually with oil. This is usually already present in the bubbler, and you should take care to keep the bubbler standing up to avoid losing oil from the bubbler.


The gas inlet will need to be sealed at this stage, as intially the setup will be purged by degassing the solvent using the pipette which has been assembled in the previous step. This can easily be achevied using a pipette teat which can be placed onto the bubbler.

Bubbler secured to the top of a condenser with a keck clip. Bubbler has a pipette teat over end which would otherwise be supplied with the nitrogen flow tube

Once both the pipette and the bubbler are in place, the setup can be used to degas the system. The precise configuration of this may be reaction dependent. For instance, air-stable solids might be added to the flask before this step, or for a setup which does not require dried solvent, the solvent could be added at this stage and degassing being carried out immediately. Otherwise, a purge step will be carried out first, followed by solvent addition, then a degassing process.

Purge

Picture of full setup of two nek round bottomed flask on hotplate. Condenser is attached to round bottom flask with a bubbler secured above it with a keck clip. A Pasteur pipette is positioned in the second neck of the two neck flask, connected to a nitrogen flow tube

With the setup as shown above, the glassware is ready to be purged with nitrogen. The nitrogen will enter the setup via the Pasteur pipette and exit the system via the oil bubbler. The nitrogen should be turned on gently, observing the bubbling rate in the oil. A gentle continuous bubbling, which doesn't dispace the oil should be observed. It is important the pressure is not increased beyond what the setup is designed to manage. The limitations are the constriction in the pipette, which only allows a low gas flow, and the gas bubbling through the oil.

An animated image of an oil bubbler on top of a reflux condenser with a steady and controlled flow of nitrogen bubbling through it. Approximately 1-2 bubbles every two seconds. The bubbler is secured to the reflux condenser with a keck clip.

The setup should be allowed to purge appropriately, then solvent can be introduced to the flask via needle and syringe. This can be acheived by drawing up solvent as detailed in the guides for dispensing liquids using a needle and syringe

Degassing solvent

The solvent can be added to the setup, e.g. by carefully lifting the condenser slightly to insert a needle into the flask. The pipette connected to the nitrogen supply may need adjustment to ensure that the pipette tip is below the level of the solvent, but not impeding the stirrer bar (note the stirrer bar can easily break the pipette tip).

Removing pipette and transferring gas supply to the oil bubbler

Upon completion of purging, the pipette is removed and replaced by a stopper. The nitrogen supply is transferred from the two-neck round bottomed flask to the oil bubbler, replacing the pipette teat.

Finishing the process