There are many different ways to prepare and run flash columns. This method describes how to ‘wet-pack’ the silica, where the silica has been 'slurried' before addition into the column. The photos show a column being prepared with 50 g of silica in a 25 mm diameter (Ø) column, being used to separate 500 mg of compound (1:100 silica ratio), collecting 20 mL fractions in test tubes using a single solvent system for elution.
There are alternative ways to prepare columns, such as 'dry packing' a column.
In order to prepare the column you will require the following items. Most of the items are generally provided collated in a tray for column chromatography, with the remaining items being provided in the lab.
A flash column for chromatography
A 250 mL conical flask to prepare the silica slurry
A 250 mL conical flask to collect solvent from the column during the preparation.
A 1 L conical flask used to prepare the column eluent
Rack of test tubes
Bellows with a take off adapter and Keck clip to secure
Round bottomed flasks to combine fractions and rotary evaporate the solvent.
Large measuring cylinder (ca. 500 mL)
Portable balances for weighing silica in a fume hood
Silica gel for chromatography
Winchesters of solvent
A tray for column chromatography.
Other lab items required to prepare a column.
In order to hold the sand/silica within the column, a cotton wool plug is fitted above the tap. The column should be securely clamped in two places around the body of the column and a 250 mL conical flask placed underneath to collect any solvent. Open the tap fully whilst loading the column.
A column securely clamped in a fume hood.
The tap end of a column showing a ball of cotton wool inserted to retain the silica.
Add around 1 cm depth of sand to the column. The sand doesn’t retain the compounds and allows anything eluting from the silica to flow into the tap immediately. The sand layer ensures that the stationary phase itself is contained only in the purely cylindrical part of the glass column; a uniform column aids resolution.
A column after an appropriate level of sand has been added.
You will need to prepare about 1 L of eluent for the loading and running of this column. The dry silica (50 g) is weighed into a conical flask (in a fumehood – take care to avoid breathing the dust).
Some of the eluent is added to the silica in order to form a slurry.
This slurry is carefully poured into the column, taking care to have a flat, level interface between the sand and silica.
Remaining silica can be washed in with more of the column eluent and silica stuck to the sides of the column can be washed down using a pipette.
Run the solvent down to the level of the silica then close the tap.
Bellows can be used to pump the solvent down the column, but must be either secured with a Keck clip, or continually held in place at all times whilst attached to the column. Before trying to attach the bellows check there isn’t silica stuck inside the ground glass joint, as this will prevent the seal being effective and may cause damage to the column.
If the bellows are clipped to the column, the pressure can be released at any point by unscrewing the plastic attachment on the take off adapter.
Once the solvent level has reached the top of the silica and the tap has been closed, the column is ready to be loaded.