Mass spectrometry is a collection of techniques used to provide mass information for molecules and fragments. There are various methods used to analyse samples using mass spectrometry, and the choice of method depends on the properties of the sample being analysed and the desired information to be obtained about a sample.
The paragraphs below provide abridged guidance for the selection of mass spectrometry methods, dependent on the sample type and desired information.
Where confirmation of the expected identity for a purified compound is sought, electrospray ionisation (ESI) should generally be the first choice of technique. ESI will result in a molecular ion peak, including isotope patterns, along with a high resolution mass. The high resolution mass result will be compared to the submitted formula and will indicate if this matches. ESI positive is suitable for a very wide array of molecules. The main exception is for organic compounds without heteroatoms (alkyl compounds).
Where a compound is known to be pure, but the identity is unknown, for example a side product isolated via column chromatography, the use of electrospray ionisation may be desirable. The results from EI can be run against large databases of known compounds where the fragmentation patterns are compared and identify possible matches.
Where a mixture of compounds requires analysis to identify the components, the use of a tandem method with chromatography is prudent. This will enable the individual components in the mixture to be separated and quantified with each component then analysed via mass spectrometry. Both gas and liquid chromatography options are available. For mixtures with compounds which are sufficiently volatile, GC-EI can be selected, with EI offering the advantage of database matching for the mixture. For less volatile mixtures, LC-## would be the method of choice.
More information on the selection of methods for mass spectrometry can be found in the following pages:
Samples will need to be prepared as detailed below for the different techniques. Please note that mass spectrometry is particularly sensitive as a technique, and it is very easy for contamination to occur. Be particularly careful handling Pasteur pipettes as contamination from pipette teats is particularly problematic. Guidance on proper Pasteur pipette use can be found on the Pasteur pipette pages.
Recording of Mass spectometry data for Teaching Labs is provided via a submission service.