To carry out a syringe filtration, you will require:
An appropriate container from which the solution can be drawn into a syringe
Syringe of appropriate size (most commonly 2 mL)
The required syringe filter which is suitable for the solvent being used and the particle size required
Collection vessel for the filtrate (often a HPLC/GC vial)
1) Transfer the solution into an apporiate container to be able to draw upliquid into a syringe (eg a sample vial or small beaker)
2) Insert the syringe into the solution to be drawn up, and pull back the plunger to draw liquid into the syringe.
Note you should draw up 0.5 mL more than the volume you require in the destination vial to account for the dead volume of the filter and any absorbtion/wetting of the membrane i.e for a chromatgrapghy vial where you require 1.5 mL of solution you should draw up 2 mL of liquid.
3) Attach the desired syringe filter onto the syringe, The filters used in teaching labs require you to screw the filters onto a Luer lock syringe (shown below) however if you are using those that can be used on a luer slip syringe they are fitted by pushing the filter firmly onto the Luer slip of the syringe.
Note that filters are usually directional, and you should ensure that the syringe is connected to the correct side of the filter.
4) The exit from filter should be placed into the collection vessel (usually a HPLC or GC vial) an the liquid gently expeled from the syringe through the filter. If you are filling a 2 mL chromatogrpahy vial this should be filled to the approproate level (usually 1.5 mL)
Care should be taken to avoid overpressurising the solution into the syringe filter. The filtration is more effective when carried out slowly. Note for Luer slip syringes, the filter should be held securely onto the syringe whilst injecting the liquid to ensure that the filter and syringe do not become detached.
5) The filtered solution should then be capped to prevent any particulate entering the solution. If you are then submitting this for HPLC or GC you should then follow the rest of the instructions on submission for these services
6) If there is any liquid remaining in your syringe this should then be returned to your original vessel for appropriate dispoal
6) The empty syringe and attached filter should be disposed of appropriately, For the Chemistry Teaching Labs the syringe and filter can be disposed together into a sharps bin.