To prepare an Mass Spec sample for submission for ESI/APCI you will need the following:
A clean 4 mL Mass Spec vial
Sample Vial Cap (optional)
A Chromatography Vial
A Chromatography vial cap with septum
MS submission number sticker
all of which are available for collection from the top of the lab at the MS Submission point, you will also need the following which is abailable from thr service beam as well as:
Clean Pasteur pipette
A micro spatula
Solvent
1. A small sample (ca. 0.5 - 1 mg) of pure compound which is free from solvent and other contaminants should be placed in to a 4 mL sample vial.
The quantity does not need to be measured accurately - the smallest quantity which can be transferred into a vial is appropriate (e.g. 1 grain of a powder or crystal or one drop of oil).
2. A Solvent for the sample to be prepared in should now be chosen, The following solvents can be used for ESI or APCI. The solvents should be attempted in the following order. If the sample does not dissolve in the first solvent attempted, try to prepare a sample using the next solvent in the list.
Ethanol
Methanol
Acetonitrile
If none of these solvents are successful, you should seek advice regarding your sample submission. Samples must not be prepared using (or contain any) DMSO or DMF as these may damage the instruments.
Both vials contain the same compound. The left is made up in Water (soluble) the right in Chloroform (insoluble)
3. Add approximately 1 mL of solvent via Pasteur pipette and use the pipette to dissolve the sample. This should yield a solution that is approximately 1 mg mL⁻¹ .
Full solvation of the product can be aided buy continuously sucking up and dispensing the solvent in the sample vial.
4. Once the sample has been dissolved, transfer one drop of this solution via Pasteur pipette into a chromatography vial, the vial should then be brought up to the 1 mL mark with the same solvent. This is approximately a 1 in 1000 dilution. An ideal concentration for ESI or APCI is ca. 1 - 10 μg mL⁻¹.
5. Ensure you have stuck the submission number sticker to the chromatography vial, so your sample can be identified by the Mass Spectrometry service technician, and complete the MS submission online form with the corresponding details to that on the label.
9. Leave the made up sample at the submission station opposite the prep room hatch, also make note of your sumbission details for locating your sample data at a later date
Mass Spectrometry is very sensitive, only use brand new clean Pasteur pipettes to take solvents from the bottle. Even tiny amounts of contamination affect Mass spec results.